Home Precision Environment Test Systems AFC 136 Filter Element Production-Control Test System
AFC 136 Filter Element Production-Control Test System product image thumbnail
AFC 136 Filter Element Production-Control Test System workflow diagram thumbnail
Topas GmbH Filter Element Production QC Systems

AFC 136 Filter Element Production-Control Test System

Automated production-control system for 100% inspection of integral separation efficiency and pressure drop on exhaust filters, HEPA cartridges and other small air-filter elements, with approximately 30-second cycles.

Test flow
40–440 m³/h; optional 1,200
Cycle time
Approximately 30 s
Specimen size
Diameter ≤ 300 mm; length ≤ 850 mm
Operation
24/7 production control
100% inspection Integral efficiency Pressure drop Barcode integration 30-second cycle
Customer-specific production criteria HEPA and exhaust-filter QC methods

Model Overview

Model System role Primary measurements Standards / application fit
AFC 136 Automated filter-element production-control system Integral separation efficiency, pressure drop and traceable pass/fail evaluation GPF/DPF, HEPA cartridges, industrial-vacuum filters and other small elements

Key Features

High-throughput inspection

Automated supply, analysis and removal enable approximately 30-second test cycles.

24/7 production design

Robust control and safety functions support continuous manufacturing environments.

Configurable acceptance criteria

Efficiency and pressure-drop limits can be set for each product family.

Minimal specimen loading

The measurement concept limits aerosol loading and helps preserve the tested product.

Identification and traceability

Barcode or QR-code integration can link results directly to production databases.

Flexible instrumentation

OPC, CPC and customer-specific aerosol-analysis devices can be selected for the application.

Technical Specifications

Parameter Unit Value / Description
TEST PERFORMANCE
Test flow rate m³/h 40–440; optional up to 1,200
Particle-size range with OPC µm 0.1–1
Particle-size range with CPC µm 0.001–1
Test cycle duration s Approximately 30
Specimen envelope Diameter ≤ 300 mm; length ≤ 850 mm
SYSTEM
Test aerosol Customer-specific, for example DEHS
Compressed-air supply bar 4–6
Duct diameter mm 150
Power supply 3 × 400 VAC; 30 A rated current
Dimensions mm 2,800 × 2,100 × 1,200

Values shown are based on the standard system configuration. Confirm the final test method, specimen fixtures, instrumentation, utilities and calibration scope before order placement.

Applications

GPF and DPF production control

Integral-efficiency and pressure-drop inspection of exhaust particle filters.

HEPA cartridge quality control

Testing of HEPA elements for vacuum cleaners and mobile air cleaners.

100% end-of-line inspection

Fast pass/fail evaluation with specimen identification and traceable data.

Small filter-element development

Comparative efficiency and pressure-drop testing for new element designs.